Commands Listed in Alphabetical Order
Syntax
DPO Models:
TRIGger:A:SETHold:DATa:SO Urce {CH1|CH2|CH3|CH4|
D0|D1|D2|D3|D4|D5|D6|D7|D 8|D9|D10|D11|D12|D13|D14|D15}
TRIGger:A:SE
THold:DATa:SOUrce?
Related Commands
TRIGger:A:SETHold:CLOCk:SOUrce
Arguments
DPO Models:
<wfm> specifies the source channel number and is one of CH1–CH4, and
D0–D15. You can specify only one waveform on a DPO.
Examples
TRIGGER:A:SETHOLD:DATA:SO URCE CH1 sets channel 1 as the clock source
for the setup and hold trigger.
TRIGGER:A:SETHOLD:DATA:SO URCE? might return
:TRIGGER:A:LOGIC:SETHOLD: DATA:SOURCE CH2 indicating that c hannel 2 is
the current clock source for the setup and hold trigger.
TRIGger:A:SETHold:DATa:THReshold
Sets or returns the data voltage threshold for setup and hold trigger. The digital
channels do not have independent trigger
levels. Channels D0–D7 share one
common trigger level and D8–D15 share another common trigger level. For
example changing the trigger level for D3 changes the levels for D0–D7.
Group
Trigger
Syntax
TRIGger:A:SETHold:DATa:TH Reshold {<NR3>|TTL|ECL}
TRIGger:A:SETHold:DATa:TH Reshold?
Arguments
TTL specifies the preset TTL high level of 1.4 V.
ECL specifies the preset ECL high level of –1.3 V.
<NR3> is the setup and hold data level, in V.
Examples
TRIGGER:A:SETHOLD:DATA:TH RESHOLD TTL specifies the preset high level of
1.4 V as the current data voltage level for
the setup and hold trigger.
TRIGGER:A:SETHOLD:DATA:TH RESHOLD? might return
:TRIGGER:A:SETHOLD:DATA:T HRESHOLD 1
.2000E+00
indicating
that 1.2 V is the current data voltage level for the setup and hold trigger.
2-336 MSO2000B, DPO2000B, M SO2000 and DPO2000 Series Oscilloscopes Programmer Manual