Acquisition
Waveform record
The instrumen
t builds the waveform record
through use of the following parameters:
Sample interval: The time between
sample points.
Record length: The number of samples
required to
ll a waveform record.
Trigger poin
t: The zero time reference in
a waveform record.
Horizontal position: When horizontal
delay is off, the horizontal position is
a percenta
ge of the waveform record
between 0 and 99.9 percent. The trigger
point and the horizontal reference are at
thesameti
me in the waveform record.
For example, if the horizontal position is
50 percent, then the trigger point is in the
middle of
the waveform record. When
horizontal delay is on, the time from the
trigger point to the horizontal reference
is the hor
izontal delay.
Interpolation
Your instrument can interpolate between the samples it acquires when it does not have all of the actual samples it needs to ll
the waveform record. Linear interpolation computes record points between actual acquired samples by using a straight line t.
Sin(x)/x interpolation computes record points using a curve t between the actual values acquired. Sin(x)/x interpolation
is the default interpolation mode because it requires fewer actual sample points than linear interpolation to accurately
represent the waveform.
MSO/DPO70000DX, MSO/DPO70000C, DPO7000C, and MSO/DPO5000B Series User Manual 39

Table of Contents

Other manuals for Tektronix DPO5204B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO5204B and is the answer not in the manual?

Tektronix DPO5204B Specifications

General IconGeneral
BrandTektronix
ModelDPO5204B
CategoryTest Equipment
LanguageEnglish

Related product manuals