NOTE
To keep users safe, always read and follow all safety warnings
provided with each of the instruments in your system.
Install the instrument
The 4200A-SCS can be used on a bench or in a rack. See
the instructions that came with your rack-mount kit if you are
installing the 4200A-SCS in a rack.
Wiring the interlock
If you need voltages greater than ±40 V for testing, you
must add an interlock switch to the xture. This ensures
that hazardous voltages are not present when the exterior
enclosure of the xture is open. It also enables the
4200A-SCS to output higher voltages when the exterior
enclosure of the xture is closed.
When the safety interlock signal is asserted (the switch is
closed and the signal is connected to +12 V), all the voltage
ranges of the SMUs are functional. When the safety interlock
signal is not asserted (the switch is open), the 200 V range
on the SMUs is disabled, limiting the nominal output to
±40 V.
If you need voltages greater than ±40 V, you must also
connect the exterior of the test xture enclosure to protective
earth (safety ground). Take care to ensure that the wiring
(FORCE, GUARD, and SENSE) in the xture does not
electrically contact the exterior of the enclosure.

Other manuals for Tektronix Keithley 4200A-SCS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley 4200A-SCS and is the answer not in the manual?

Tektronix Keithley 4200A-SCS Specifications

General IconGeneral
BrandTektronix
ModelKeithley 4200A-SCS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals