Built-in security features
When to use TekSecure
Use the TekSecure function to erase setup and reference waveform data stored in
internal flash memory. Using TekSecure will not affect the calibration of the
instrument because the calibration constants are stored on the Acquisition board,
completely separate from any acquisition data. This allows complete erasure or
removal of any secure data without affecting the oscilloscope calibration. It also
allows the oscilloscope to be calibrated in a non-secure site and then used in a
secure site without the need for recalibration.
The 3 Series MDO oscilloscopes has USB host ports: located at the lower right
corner of the front panel, and on the rear panel. Any USB flash devices can be
removed and stored or destroyed.
The TekSecure function does the following:
Replaces all waveforms in all reference memories with null sample values
Replaces the current front-panel setup and all stored setups with the default
setup values
Calculates the checksums of all reference waveform memory and setup
memory locations to verify successful completion of waveform and setup
erasure
Displays a dialog indicating whether the secure erase was successful or
unsuccessful
NOTE. TekSecure does not erase or change factory calibration constants or
Ethernet settings.
3 Series MDO Declassification and Security Instructions 11

Other manuals for Tektronix MDO34

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO34 and is the answer not in the manual?

Tektronix MDO34 Specifications

General IconGeneral
BrandTektronix
ModelMDO34
CategoryTest Equipment
LanguageEnglish

Related product manuals