Appendix E: Search and Trigger Command Sequence
Examples
The following are some example command sequences that show a variety of
different sear ches and triggers. The commands in these sequence s are not
order-dependent.
To use these examples, connect channel 1, channel 2, channel 3 and channel 4
(if available) to the probe compensation signal located on the right hand side
of the front panel.
The search and trigger command group sections contain
more information on
general search and trigger concepts.
Example 1: Single Threshold Edge Search
Goal: Search the channel 2 waveform and place a mark at each instance where it
crosses below a threshold of 1.4 volts.
Command Comment
*RST Resets the oscilloscope. Wait for the reset to complete
(approximately 3 seconds).
:SELect:CH2 1 Turns the CH2 waveform on.
:AUTOSet EXECute Autosets the displayed waveform CH2. Wait for the autoset to
complete (approximately 3 seconds).
:SEARCH:SEARCH1:TRIGger:A:TYPe EDGE Species that this will be an edge search (a mark will be placed
when the source waveform passes through a specied threshold
level in the specied direction).
:SEARCH:SEARCH1:TRIGger:A:EDGE:SOUrce CH2 Species the CH2 waveform as the source waveform.
:SEARCH:SEARCH1:TRIGger:A:LOWerthreshold:CH2 TTL Species 1.4 volts as the threshold level.
:SEARCH:SEARCH1:TRIGger:A:EDG E:SLOpe FALL Species the falling edge as the direction.
:SEARCH:SEARCH1:STATE 1
Turns the search on.
:SEARCH:SEARCH1:TOTal? Returns 4, indicating that CH2 fell below the 1.4 volt threshold 4
times.
Example 2: Single Threshold Edge Trigger
Goal: Trigger on the channel 2 waveform when the waveform crosses below a
threshold of 1.4 volts.
Command Comment
*RST Resets the oscilloscope. Wait for the reset to complete (approximately
3 seconds).
:SELect:CH2 1 Turns the CH2 waveform on.
MSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual E-1

Other manuals for Tektronix MSO2012B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO2012B and is the answer not in the manual?

Tektronix MSO2012B Specifications

General IconGeneral
BrandTektronix
ModelMSO2012B
CategoryTest Equipment
LanguageEnglish

Related product manuals