Commands Listed in Alphabetical Order
Examples
DATA:ENCDG? mi
ght return
:DATa:ENCDG SRPBINARY for the format of the
outgoing w aveform data .
DATA:ENCDG RP
Binary
sets the data encoding format to be a positive integer
where the most significant byte is transferred first.
DATa:RESOlution
Sets or returns whether the CURVe? query returns full resolution records
(acquired data) or reduced resolution records (filtered/displayed d ata).
NOTE. For more information on querying waveform data, see Appendix B:
Waveform
Data in MSO/DPO2000B and MSO/DPO2000 Series Instruments.)
Group
Waveform Transfer
Syntax
DATa:RESOlution {FULL|REDUc
ed}
DATa:RESOlution?
Arguments
FULL sets the instrument to return the full undecimated record acquired by the
instrument. The full resolution records are not subject to the effects of FilterVu.
Full resolution record lengths are 100,000, 125,000, 1,000,000 or 1,250,000
points.
REDUced sets the instrument to return short waveforms that are decimated from
the full acquisition record. Reduced records are subject to the low pass filtering
effects of FilterVu settings. Reduced record lengths range from 800 to 6250
points, depending o n the acquisition settings.
Examples
DATA:RESOLUTION? might return :DATA:RESOLUTION: FULL indicating that
the CURVe? query returns full resolution records (acquired data).
DATA:RESOLUTIONREDUCED sets the CURVe? returns to reduced resolution
records.
DATa:SOUrce
Sets or returns the location of the waveform data transferred from the oscilloscope
by the CURVe?query.
NOTE. For more information on querying waveform data, see Appendix B:
Waveform Data in MSO/DPO2000B and MSO/DPO2000 Series Instruments.)