Display Waveform Data
6. To change the scale (height) of the d igital
channels, push the lower-menu Height button.
NOTE. The S (Small) selection will display
each waveform at 0.2 divisions tall. The M
(Medium) selection will display each waveform
at 0.5 divisions tall. The L (Large) selection will
display each waveform at 1 division tall. L only
works if there is enough room in the display to
display the waveforms. You can display up to 10 L
waveforms at one time.
7. You can label individual digital c hanne ls for
easier identication. (See page 84, Labeling
Channels and Buses .)
8. To g roup some or all of the digital channels
together, m ove the channels right next to each
other. All the channels that are next to each
other automatically form a group .
Yo
u can see the g roups by pressing
the s
ide-bezel Select item a nd turning
multip
urpose knob a.
When a gro
up is selected, turn m ultipurpose
knob b to move t
he whole group.
184 MSO4000 and DPO4000 Series Oscilloscopes User Manual

Table of Contents

Other manuals for Tektronix MSO4032

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO4032 and is the answer not in the manual?

Tektronix MSO4032 Specifications

General IconGeneral
BrandTektronix
ModelMSO4032
CategoryTest Equipment
LanguageEnglish

Related product manuals