Specications
Table 2: Digita
l channel input specications, MSO4000 only
Characteristic Description
Threshold vol
tage range
-2Vto+5V
Digital threshold accuracy
±[100 mV + 3% o
f the threshold setting after calibration]
Timing resolution
2 ns for the m ain memory and 60.6 ps for MagniVu memory
Table 3: Horizontal and acquisition system specications
Characteristic
Description
Long-term sample rate and delay
time accuracy
±5 ppm over any 1mstimeinterval
Seconds/Division range
Record Length
DPO4104, MSO4104
DPO4054, DPO 4034, DPO4032,
MSO4054, MSO4034, MSO4032
1 K 400 ps 40 s 1 ns 40 s
10 K 400 ps 400 s 1 ns 400 s
100 K
1M
10 M
400 ps 1,000 s 1 ns 1,000 s
1-2-4 sequence 1-2-4 sequence except one position in each
record length selection, which is 0.8-2-4.
Minimum pulse widthPeak Detect or Envelope mode pulse
response, typical
DPO4104, MSO4104: > 200 ps
DPO4054, DPO4034, DPO4032, MSO4054, MSO4034, MSO4032: > 400 ps
Sample-rate range DPO4104, MSO4104: 5 GS/s-0.1 S/s
DPO4054, DPO4034, DPO4032, MSO4054, MSO4034, MSO4032: 2.5 G S/s 0.1 S/s
Record length range 10 M, 1 M, 100 k, 10 k, 1 k
Maximum update rate
Maximum triggered acquisition rate: 50,000 wfm/s
Aperture Uncertainty, typical
(3 ps + 0.1 ppm × record duration)
RMS
, for records having duration 1minute
Number of Waveforms for Average
Acquisition Mode
2 to 512 waveforms
Default of 16 waveforms
Table 4: Trigger specications
Characteristic Description
Aux In (E xternal) trigger maximum
input voltage
The maximum input voltage at the BNC, between center conductor and shield, is 400 V
peak
(DF 39.2%), 250 V
RMS
to 2 MHz derated to 5 V
RMS
@ 500 MHz.
The maximum transient withstand voltage is ±800 V
peak
.
Aux In (External) trigger input
impedance, typical
1M ±1% in parallel with 13 pF ±2 pF
Aux In (External) trigger bandwidth,
typical
250 MHz ±20%
Trigger bandwidth, Edge, Pulse, and
Logic, typical
DPO4104, MSO 4104: 1 G Hz
DPO4054, MSO 4054: 500 MHz
DPO4034, DPO4032, MSO4034, MSO4032: 350 MHz
Time range Accuracy
1nsto500ns
±(20% of setting + 0.5 ns)
Time accuracy for Pulse, Glitch,
Timeout, or Width triggering
520 ns to 1 s
±(0.01% of setting + 100 ns)
MSO4000 and DPO4000 Series Specications and Performance Verication 7

Other manuals for Tektronix MSO4054

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO4054 and is the answer not in the manual?

Tektronix MSO4054 Specifications

General IconGeneral
BrandTektronix
ModelMSO4054
CategoryTest Equipment
LanguageEnglish

Related product manuals