Item Inspect for Repair action
Capacitors Damaged or leaking cases.
Corroded solder on leads or
terminals.
Remove and replace damaged circuit board.
Wiring and cables Loose plugs or connectors.
Burned, broken, or frayed wiring.
Firmly seat connectors. Repair or replace modules with defective
wires or cables.
Chassis Dents, deformations, and
damaged hardware.
Straighten, repair, or replace defective hardware.
Cleaning procedure interior
To clean the oscilloscope interior, perform the following steps:
1. Blow off dust with dry, low-pressure, deionized air (approximately 9 psi).
2. Remove any remaining dust with a lint-free cloth dampened in isopropyl
alcohol (75% solution) and then wipe with a lint-free cloth dampened with
warm deionized water. A cotton-tipped applicator is useful for cleaning in
narrow spaces and on circuit boards.
STOP. If, after performing steps 1 and 2, a module is clean on inspection,
skip steps 3, 4, and 5. Proceed with step 6.
3. If steps 1 and 2 do not remove all the dust or dirt, all modules in the
oscilloscope except the Display Module and Front Panel Module may be
spray washed using a solution of 75% isopropyl alcohol.
■
During cleaning of the Acquisition board (Main board), make sure that
no residue is left in the attenuators. Debris or residue can cause several
problems with signal fidelity that cannot be corrected by calibration.
When handling the board, avoid handling the attenuators since skin oils
may cause problems.
■
Extensive cleaning of the Display Module is difficult because water must
not be trapped between the display shield and the display. The polarizers
integrated into the display degrade with long-term exposure to water. The
display shield and display must be replaced as an assembly if significant
moisture or debris become trapped inside. Cotton tipped applicators and
lint-free rags moistened in isopropyl alcohol and/or deionized water offer
the best cleaning with the least opportunity for damage to the assembly.
■
The Front-Panel module can be damaged by deposition of film or debris
inside the knob encoders. To avoid this, be careful not to drip liquid or
blow debris into the encoders.
4. Spray wash dirty parts with the 75% isopropyl alcohol solution and wait
60 seconds for the majority of the alcohol to evaporate.
5. Use hot (49 °C to 60 °C / 120 °F to 140 °F) deionized water to thoroughly
rinse the parts.
Maintenance
4 TBS1000C Series Oscilloscope Service Manual