EasyManuals Logo
Home>Tektronix>Test Equipment>TBS1052C

Tektronix TBS1052C User Manual

Tektronix TBS1052C
40 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #28 background imageLoading...
Page #28 background image
Push menu button Select menu option Select setting
Default setup - -
Channel 1 Probe 1X
2. Connect the oscilloscope to the time mark generator as shown in the following figure:
3. Set the time mark generator period to 10 ms.
4. Set the oscilloscope Vertical Scale (volts/division) to 500 mV/div.
5. Set the oscilloscope Main Horizontal Scale (seconds/division) to 1 ms/div.
6. Push the Trigger Level knob to activate the Set To 50% feature.
7. Use the Vertical Position control to center the test signal on screen.
8. Use the Horizontal Position control to set the position to 10.00 ms.
9. Set the oscilloscope Horizontal Scale (seconds/division) to 200 ns/div.
10. Check that the rising edge of the marker crosses the center horizontal graticule line within ±2 divisions of the vertical center
graticule line, as shown in the following figure:
NOTE. One division of displacement from graticule center corresponds to a 25 ppm time base error.
11. Disconnect the test setup.
Performance verification
18 TBS1000C Series Specification and Performance Verification

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TBS1052C and is the answer not in the manual?

Tektronix TBS1052C Specifications

General IconGeneral
BrandTektronix
ModelTBS1052C
CategoryTest Equipment
LanguageEnglish

Related product manuals