Performance Tests
TDS Family Performance Verification and Specifications
1–23
TDS 500B/700A Test Record
Photocopy this and the next three pages and use them to record the performance
test results for your TDS 500B/700A.
TDS 500B/700A Test Record
Instrument Serial Number: Certificate Number:
Temperature: RH %:
Date of Calibration: Technician:
TDS 500B/700A Performance Test Minimum Incoming Outgoing Maximum
Offset Accuracy
CH1 Offset +1 mV
+101 mV
+1.01 V
– 1.6 mV
– 25.1 mV
– 251 mV
__________
__________
__________
__________
__________
__________
+ 1.6 mV
+ 25.1 mV
+ 251 mV
CH2 Offset +1 mV
+101 mV
+1.01 V
– 1.6 mV
– 25.1 mV
– 251 mV
__________
__________
__________
__________
__________
__________
+ 1.6 mV
+ 25.1 mV
+ 251 mV
CH3 or AX1 Offset +1 mV
+101 mV
+1.01 V
– 1.6 mV
– 25.1 mV
– 251 mV
__________
__________
__________
__________
__________
__________
+ 1.6 mV
+ 25.1 mV
+ 251 mV
CH4 or AX2 Offset +1 mV
+101 mV
+1.01 V
– 1.6 mV
– 25.1 mV
– 251 mV
__________
__________
__________
__________
__________
__________
+ 1.6 mV
+ 25.1 mV
+ 251 mV
DC Voltage Measurement Accuracy (Averaged)
CH1 5 mV Vert scale setting,
–5 Div position setting, +1 V offset
+ 1.0355 V __________ __________ + 1.0445 V
CH1 5 mV Vert scale setting,
+5 Div position setting, –1 V offset
– 1.0445 V __________ __________ – 1.0355 V
CH1 200 mV Vert scale setting,
–5 Div position setting, +10 V offset
+ 11.5385 V __________ __________ + 11.6615 V
CH1 200 mV Vert scale setting,
+5 Div position setting, –10 V offset
– 11.6615 V __________ __________ – 11.5385 V
CH1 1 V Vert scale setting,
–5 Div position setting, +10 V offset
+ 17.7785 V __________ __________ + 18.2215 V
CH1 1 V Vert scale setting,
+5 Div position setting, –10 V offset
– 18.2215 V __________ __________ – 17.7785 V
CH2 5 mV Vert scale setting,
–5 Div position setting, +1 V offset
+ 1.0355 V __________ __________ + 1.0445V
CH2 5 mV Vert scale setting,
+5 Div position setting, –1 V offset
– 1.0445 V __________ __________ – 1.0355 V
CH2 200 mV Vert scale setting,
–5 Div position setting, +10 V offset
+ 11.5385 V __________ __________ + 11.6615 V

Other manuals for Tektronix TDS 684A