Maintenance
6–32
TDS3000 Series Service Manual
3. Place the oscilloscope face down on a soft surface (such as an anti-static
mat), with the bottom facing you.
4. Disconnect the bezel button flex circuit from J200 by gently pulling the flex
circuit out of the connector. Gently remove the flex circuit from the front
panel and place on a clean protected surface.
5. Remove the front-panel board by pulling back board snaps and lifting the
lower right corner of the front-panel board. Pull back the board snaps in
numeric order, starting with snap one, as shown in Figure 6–21.
Board snap (1)
Board snap (3)
Board snap (2)
Lift board from
this corner
Board snap (4)
Board snap (5)
Board snap (6)
Figure 6–21: Remove front-panel board