Preface
This document describes how to clear or s anitize the data storage (memory)
devices in the products listed below. The instructions also describe how to
declassify a
n instrument that is not functioning. These procedures are useful for
customers with data security concerns.
Products
This document applies to the following components of the TG8000 Multiformat
Test Signal Generator:
TG8000 mainframe
GPS7 GPS Synchronization Timecode module
HD3G7 HD 3 Gb/s SDI Video Generator module
SDI7 Dual Channel SD/HD/3G SDI Video Generator module
Required d ocuments
To perform the procedures in this document, you will need to have access to the
TG8000 manuals listed b
elow. These manuals are available on the Tektronix Web
site at www.tektronix.com/downloads.
TG8000 Multiformat Test Signal Generator User Manual
(Tektronix part number 071-3036-XX; 071-3037-XX, Japanese;
071-03038-XX, Russian)
TG8000 Multiformat Test Signal Generator Service Manual
(Tektronix part number 077-0687-XX)
Terms
The following terms may be used in this document:
Clear. This removes data on media/memory before reusing it in a secured
area. All reusable memory is cleared to deny access to previously stored
information by standard means of access.
Erase. This is equivalent to clear.
Media storage/data export device. Any of several devices that are used to
store or export data from the instrument, such as a USB port.
Nonvolatile memory. Data is retained when the instrument is powered off.
Power off. Some instruments h ave a “Standby” mode, in which power is
still supplied to the instrument. For clearing data, putting the instrument in
Standby mode does not qualify as powering off. For these products, you must
either push a rear-panel OFF switch or remove the power source from the
instrument.
Remove. This is a p hysical means to clear the data by removing the memory
device from the instrument. Instructions are a vailable in the product Service
Manual.
TG8000 Declassification and Security Instructions iii