Set the attenuation coefficient in the probe menu and the switch on the probe to 10x, and connect the
probe to CH1. Make sure the probe’s connector is properly connected with the oscilloscope. Connect the
probe’s main clip and ground clip to the oscilloscope’s compensating signal and ground terminal
respectively. Open CH1 and press the AUTO button.
View displayed waveforms
Excessive Compensation Correct Compensation Insufficient Compensation
Figure 2 Compensating Calibration of Probe
If the displayed waveform is look like the above “Insufficient Compensation” or “Excessive
Compensation”, use a non-metallic screwdriver to adjust the probe’s variable capacitance until the display
matches the "Correct compensation" waveform.
Warning: To avoid electric shock when measuring high voltage using the probe, please ensure
that the probe insulation is in good condition and avoid physical contact with any metallic part of
the probe.