Model UT106: OPERATING MANUAL
B. Replacing the Fuses (see figure 24)
Warning
To avoid electrical shock or arc blast, or personal
injury or damage to the Meter, use specified fuses
ONLY in accordance with the following procedure.
Screw
Holster
(figure 24)
To replace the Meter’s fuse:
Turn the Meter off and remove all connections from
the terminals.
Remove the holster from the Meter.
Remove the 3 screws from the case bottom, and
separate the case top from the case bottom.
Remove the fuse by gently prying one end loose, then
take out the fuse from its bracket.
Install ONLY replacement fuses with the identical type
and specification as follows and make sure the fuse
is fixed firmly in the bracket.
Fuse 1: 10A, 250V, fast type, 5x20 mm.
Fuse 2: 315mA, 250V, fast type, 5x20 mm.
Rejoin the case bottom and case top, and reinstall the
3 screws and holster.
1.
2.
3.
4.
5.
6.
Replacement of the fuses is seldom required. Burning
of a fuse always results from improper operation.