Model UT107: OPERATING MANUAL
H. Temperature Measurement (see figure 9)
( figure 9)
Warning
To avoid harm to yourselves, never attempt to input
an effective voltage over 60V in DC or 30V in AC.
The temperature measurement range is -40
0
C ~ 1000
0
C.
Insert the red test lead into the mA
0
C terminal
and the black test lead into the COM terminal.
Set the rotary switch to
0
C.
Place the temperature probe on the outside or
internal of the object being measured.
The measured value shows on the display, the unit
is
0
C.
1.
2.
3.
4.
Black Red
Please choose a correct temperature probe. The
included temperature probe can only measure up to
250
0
C. You need to select another temperature probe
for any measurement higher than 250
0
C.
The Meter display ā€œ1ā€ when there is no signal input.
The Meter display its internal nearest temperature
value when the two input terminals are shorted.
When temperature testing has been completed,
disconnect the connection between the temperature
probe and the circuit under test.
l
l
l
l
To measure temperature, connect the Meter as below: