47
Model UT71A/B/C/D/E: OPERATING MANUAL
Press STORE once, STO and "no.xxxx" appears to
confirm the operation and the left secondary display
shows the current measurement reading. Press
to toggle between clearing the stored readings and
start from the first readings or start from the last stored
reading. Right secondary display shows the original
number of records.
Press STORE the second time, STO appears. The
left secondary display shows the storing time interval
in second, it is preset to zero. To change the interval
in second by pressing + or - button. The interval can
l
l
l
l
l
l
be as high as 255 seconds or as low as 0 second.
Press and hold STORE to access the quick setting.
Press STORE the third time, STO and no appears.
The left secondary display shows the index number
increase one. The right secondary display shows
the value of the corresponding index number, the
primary display shows the current measurement
reading.
If there is not set time to store the reading, each press
of STORE to store one reading. An index number
increase one.
The maximum number of stored reading is 100 (for
UT71C and UT71E) and 9999 (for UT71D). When
the stored readings memory is full, the Meter will stop
storing data.
To exit, press EXIT.
Automatic power off feature will be disabled after
entering this mode.
Chapter 4
Using Store, Recall & Send Features
Introduction
Chapter 4 shows you how to use stores, recall and
communication features available on the Meter
Storing and Clearing Readings
To store readings, proceed as follows: