62
Model UT71A/B/C/D/E: OPERATING MANUAL
Remarks:
l
l
l
Input Impedance: Approx 10M .
Overload Protection: 1000V.
Display:
a) True rms are valid from 10% of range to 100% of range
b) AC crest factor can be up to 3.0 except 1000V where it is 1.5.
c) A residual reading of 80 digits with test leads shorted, will not affect stated accuracy.
d) When frequency is lower than 100kHz, the accuracy guarantee range 10%-100%.
e) When making AC+DC measurment, the accuray need to add (1%+ 35 digits) of reading based on the above table.