14 7
Measured (continued):
Reportable Reference Range
Test Units Range (arterial) (venous)
pH 6.50 – 8.20 7.35 – 7.45 7.31 – 7.41
PCO
2
mmHg 5 – 130 35 – 45 41 – 51
kPa 0.67 – 17.33 4.67 – 6.00 5.47 – 6.80
TCO
2
mmol/L (mEq/L) 5-50 23 – 27 24 – 29
(on the CHEM8+ cartridge only)
PO
2
mmHg 5 – 800 80 – 105
kPa 0.7 – 106.6 10.7 – 14.0
Ionized Calcium/iCa mmol/L 0.25 – 2.50 1.12 – 1.32 1.12 – 1.32
mg/dL 1.0 – 10.0 4.5 – 5.3 4.5 – 5.3
Urea Nitrogen/BUN mg/dL 3 – 140 8 – 26 8 – 26
Urea mmol/L 1 – 50 2.9 – 9.4 2.9 – 9.4
mg/dL 6 – 300 17 – 56 17 – 56
Hematocrit/Hct %PCV 15 – 75 38 – 51 38 – 51
Fraction 0.15 – 0.75 0.38 – 0.51 0.38 – 0.51
Celite
®
Activated seconds 50 – 1000 74 – 125 (Prewarm) 74– 125 (Prewarm)
Clotting Time/ 84 – 139 (Nonwarm) 84 – 139 (Nonwarm)
Celite
ACT
The range from 80 - 1000 seconds has been verified through method comparison studies.
Kaolin Activated seconds 50 – 1000 74 – 137 (Prewarm) 74 – 137 (Prewarm)
Clotting Time/ 82 – 152 (Nonwarm) 82 – 152 (Nonwarm)
Kaolin
ACT
The range from 77 - 1000 seconds has been verified through method comparison studies.
Prothrombin Time/PT INR 0.9 – 8.0
Performance characteristics have not been established for INRs above 6.0.
Troponin I / cTnI ng/mL (µg/L) 0.00 – 50.0 0.00 – 0.03*
0.00 – 0.08**
Performance characteristics have not been established for cTnI values above 35.00 ng/mL.
* Represents the 0 to 97.5% range of results.
** Represents the 0 to 99% range of results.
Creatinine Kinase MB/ ng/mL (µg/L) 0.0 – 150.0 0.0 – 3.5***
CK-MB
***Represents the 0 to 95% range of results.
β-Type Natriuretic pg/mL (ng/L) 15 – 5000 <15 – 50#
Peptide / BNP
# Represents the 0 to 95% range of results.
Total Beta-Human IU/L 5.0 – 2000.0 <5.0
Chorionic Gonadotropin
/ß-hCG
Celite is a registered trademark of Celite Corporation, Santa Barbara, CA. for its diatomaceous earth products.
Procedure
1. Turn the handheld on and press to access the Administration Menu.
2. Press for Quality Tests.
3. Press for Control.
4. Press for i-STAT Cartridge.
5. Scan or enter Operator ID.
6. Enter the control lot number.
7. Enter the cartridge lot number.
8. Fill a cartridge with the control and close the cover.
9. Insert the cartridge into the cartridge port.
10. Enter chart page information if applicable.
11. View results on handheld’s display.
12.
Remove and discard cartridge when Cartridge Locked message disappears.
13. Press for Test Options on the results page and press for Next Level if
testing another level of control.
3
1
1
1
1
MENU
Verification of Cartridges Using 2 Levels of Specified Controls
Cartridge Type Control Type
G, Crea, G3+, CG4+
TriControls
or
i-STAT controls
CHEM8+, E3+, EC4+, 6+
TriControls
EC8+, EG6+, EG7+, CG8+
ACTk, ACTc i-STAT ACT controls
PT/INR i-STAT PT/INR controls
cTnI i-STAT cTnI controls
CK-MB i-STAT CK-MB controls
BNP i-STAT BNP controls
ß-hCG i-STAT ß-hCG control
n Verify that the storage conditions listed above have been maintained.

Other manuals for Abbott i-STAT

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Abbott i-STAT and is the answer not in the manual?

Abbott i-STAT Specifications

General IconGeneral
BrandAbbott
Modeli-STAT
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals