Introduction
Chapter 18
NOTE The 11708A 30 dB attenuator is intended for use only at the 1 mW, 50 MHz power
reference of the power meter. Its function as a calibration reference may be
compromised if used for other purposes.
26.5 GHz and 33 GHz Frequency operation (8485A, 8485A-033, 8485D and
8485D-033)
8480 series sensors that operate up to 26.5 GHz and 33 GHz are fitted with
APC-3.5mm (m) connectors as standard. To convert the APC-3.5mm (m) connector
for calibration an adapter (APC-3.5 (f) to Type-N (m)) is included with the power
sensors. Figure 1-4 shows the parts included with your power sensor.
NOTE The APC-3.5mm to Type-N adapter is intended for use only at the 1 mW, 50 MHz
power reference of the power meter. Its function as a calibration reference may be
compromised if used for other purposes.
Table 1-2 11708A 30 dB attenuator characteristics
Characteristic Limits Comments
11708A accuracy at
50 MHz, 25 °C
30 ± 0.05 dB Accuracy traceable to National Institute of
Standards and Technology (NIST), with a
temperature coefficient typically 0.003 dB per
°C.
Dimensions Length: 60 mm (2.4 in)
Diameter: 20 mm (0.8 in)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8481A and is the answer not in the manual?

Agilent Technologies 8481A Specifications

General IconGeneral
BrandAgilent Technologies
Model8481A
CategoryAccessories
LanguageEnglish

Related product manuals