Introduction 1
Agilent 8494/95/96G/H Attenuators Operating and Service Manual 17
Maximum SWR
Maximum Residual Attenuation
Attenuation Repeatability
±0.03 dB max (5 million cycles per section).
Table 7 Maximum SWR
Instrument Frequency Range (GHz) Maximum SWR
8495G dc to 4 1.35
8495H dc to 8
8 to 12.4
12.4 to 18
1.35
1.5
1.7
8494G,
8496G
dc to 4 1.5
8494H,
8496H
dc to 8
8 to 12.4
12.4 to 18
1.5
1.6
1.9
Table 8 Maximum Residual Attenuation
Instrument Maximum Residual Attenuation
8494G,
8494H
0.6 dB + 0.09 dB/GHz
8495G,
8495H
0.4 dB + 0.07 dB/GHz
8496G,
8496H
0.6 dB + 0.09 dB/GHz