50 Chapter 2
cdma2000/IS-2000 Mobile Test Application
How Do I Perform a Handoff?
How Do I Perform a Handoff?
1. On the Call Setup screen, press the Call/Handoff Setup (F5) key.
2. Highlight the Handoff System Type parameter and select the desired system.
3. Highlight the Handoff Cell Band parameter and select the desired band.
4. Highlight the Handoff Channel parameter and select the desired channel.
5. Press the Execute Handoff
(F5) key to complete the handoff, or the Close Menu (F6) key
to abort the handoff.
6. Check for Connected in the Active Cell status field.