16
Verify Proper Installation of Option 1E1
1. Press: Utility > Instrument Info > Help Info > Options Info.
2. Make sure Option 1E1 is listed.
NOTE If you enable an option that does not have the required hardware installed, the menus for
that option will be activated but the option cannot operate.
Re-Assembling the Instrument
Refer to Figure 1 and Figure 2.
1. Reinstall the inner and outer instrument covers by reversing the order for removal.
2. Torque all T-10 screws to 9 in-lbs.
3. Torque all T-15 and T-20 screws to 21 in-lbs.
4. Cut out an Option 1E1 label and attach it to the rear panel, near the original serial number label.
Verify Signal Generator Calibration
1. Turn the signal generator on and allow it to warm up for 30 minutes.
2. Perform the following adjustment and performance tests, in the order listed, to verify signal generator
calibration. Refer to the PSG Family Signal Generators Service Software.
•Self Tests
Power Flatness Calibration
Lowband Attenuator Calibration (<3.2GHz)
High Band Attenuator Calibration (>3.2GHz)
V-Band Attenuator Calibration (>50GHz)
Maximum Leveled Output Power
Power Level Accuracy

Other manuals for Agilent Technologies E8257D

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E8257D and is the answer not in the manual?

Agilent Technologies E8257D Specifications

General IconGeneral
BrandAgilent Technologies
ModelE8257D
CategoryInverter
LanguageEnglish

Related product manuals