46 U1241B/U1242B User’s and Service Guide
5 Performance Tests and Calibration
Calibration Security
The calibration security code prevents accidental or unauthorized
adjustments to the meter. The meter is secured when the meter is shipped
from factory. Before performing any adjustment to the meter, you are
required to unsecure the meter by entering the correct security code (see
Unsecuring the Meter for Calibration). The security code may contain up
to 4 numeric characters.
Unsecuring the Meter for Calibration
Before performing adjustment to the meter, you are required to unsecure
the meter by entering the correct security code. The security code is set to
1234 when the meter is shipped from the factory. The security code is
stored in non-volatile memory, and does not change when power is off.
Unsecuring the Meter from the Front Panel
1 Turn the rotary switch to .
2 Press and simultaneously to enter the Calibration
Security Code entry mode. The primary display shows 5555 and the
secondary display shows SEU.
3 Press or to step each character in the code. Press
or to change the value of the selected character.
4 Press (Save) when done.
5 If the correct security code is entered, the secondary display will show
“PAS ”. If an invalid code is entered, the meter will shows error code
“E02” on the secondary display for approximately 3 seconds and return
to the Calibration Security entry mode.
You can unsecure and change the security code from the meter front panel. If you forget
your security code, see Unsecuring the Meter Without the Security Code.