EasyManuals Logo
Home>Agilent Technologies>Multimeter>U1241B

Agilent Technologies U1241B User's And Service Guide

Agilent Technologies U1241B
81 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #62 background imageLoading...
Page #62 background image
48 U1241B/U1242B User’s and Service Guide
5 Performance Tests and Calibration
Using the Front Panel for Adjustments
This section describes the process used to perform adjustments from the
front panel.
Selecting the Adjustment Mode
Unsecure the meter, see Unsecuring the Meter for Calibration or
Unsecuring the Meter Without the Security Code. Once the meter has been
unsecured, the reference value will be indicated on the primary display.
Entering Adjustment Values
1 Press and to step through each character in the primary
display.
2 Press and to change the value of the corresponding
character from digits 0 through 9.
3 Apply the respective corresponding input signal from the recommended
test equipment in Table 5- 1.
4 Press (Save) when done to start calibration.
NOTE
Ensure that the accuracy of the input signal adheres to the Valid Adjustment Input Values
in Tabl e 5-3.

Table of Contents

Other manuals for Agilent Technologies U1241B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies U1241B and is the answer not in the manual?

Agilent Technologies U1241B Specifications

General IconGeneral
BrandAgilent Technologies
ModelU1241B
CategoryMultimeter
LanguageEnglish

Related product manuals