R0368
Cod.R0368‐rev.1.0
(03/2020)
Page 10/14
Startingfromthehomepage,presstheF2keyto
entertheServiceMenu.
Figure17
UsetheF2/F3keystoselectthe"CalibrationMenu"
option,thenconfirmwithF4.
Accessthroughthefollowingpassword:
“F8;F7;F6;F8”.
Figure18
Select, using the F3/F2 keys, the "Detector
Calibration"optionandconfirmwiththeF4key.
Figure19
Ifrequired,pressF3tobeabletoentertheserial
numbers of the detectors so that they can
eventuallybementionedintheprintreport.
Ifnecessary,alsoentertheserialnumberofthetool
andthenameoftheoperator.
ContinuebypressingF4.
Figure20