Trouble shooting
by Blum-Novotest 2012 33
Error LED Status after battery change
LED Status flashes red
Control sends signal „deflected“ without deflecting the stylus
No pressure compensation after transport by airfreight
Provide pressure compensation:
1. Remove cover from battery box
2. Loosen thread pin (3a), approx. 2 revolutions
3. Deflect stylus several times in Z-direction, wait for approx. 1 minute
4. Then tighten thread pin again
5. Fit battery box cover precisely and mount it
Error No activation
No activation of the probe is possible
(1) Battery is empty
(2) Probe is outside of the transmission range
(3) No pairing of the probe on IF59-A2
(1) Insert new batteries
(2) Check transmission range between receiver and probe
Check adjustment of transmission power (probe and IF59-A2)
(3) Check pairing