Troubleshooting
Problem Reason Solution
Heart rate
symbol
Appears in the measurement
condition and ashes when
pulse is detected
• Measurement in progress, remain quiet.
Low battery
indicator
Appears when the battery
voltage is excessively low or the
positions of the batteries are
incorrect
• Replace all four batteries with new ones.
Insert the batteries in the correct
positions. Be aware of +/- positions.
Measuring
error
Appears when the accurate
blood pressure and pulse could
not be obtained
• Press “start/stop” button again and
remeasure.
• Check if cu is wrapped according
to instructions.
• Check that there are no kinks in tube.
• Check that hand is relaxed.
• Check if talking or moving
during measurement.
• Check if posture is correct.
E0 Measuring error • The pulse signals on the cu are too
weak. Re-position the cu and repeat the
measurement.
E1 Pumping error • The cu is not fastened properly. Re-position
the cu and repeat the measurement.
E2 Out of Range Results Error • Relax for a while and repeat the
measurement. Make sure the cu is correctly
positioned on your arm.
E3 Pump time error • The cu is not fastened properly. Re-t the
cu, and take a measurement again.
E254 Device internal error • Remove and then reinsert the batteries. If the
problem persists, contact customer support.
E255 Calibration error • Remove and then reinsert the batteries. If the
problem persists, contact customer support .
E256 An error occurred during the
Bluetooth Firmware update
• Device will automatically restart using the
original Bluetooth rmware. Re-initiate the
rmware download process or follow the on-
screen instruction from Braun APP.
English
Specications
Method of measurement Oscillometric
Model number BUA6350
Range of measurement
Rated cu pressure: 0mmHg~300mmHg
Measurement pressure: SYS: 60~255mmHg DIA: 40~200mmHg
Pulse value: 40-199 beat/minute
Accuracy Pressure +/- 3 mmHg
Pulse +/- 5 % Max.
Ination Deluxe automatic
Display LCD with backlight
Sets of device memory 60 sets per user
Sets of Bluetooth memory 200 sets per user
Cu size Small cu = 22-32cm arm circumference
Large cu = 32-42cm arm circumference
Operating condition +10 °C ~ +40 °C, less than 85 % R.H.
Storage condition -20 °C ~ +60 °C, less than 85 % R.H.
Operating atmospheric pressure 860-1060 hPa
Unit weight Approximately 500gr (without batteries)
Power supply Alkaline battery: 4 x AA (LR6) 1.5V
Battery life 300 times measurement
Auto power o Whenever not used for 1 minute
Accessories 4 batteries, 2 arm cus with tube, instruction manual,
quick reference guide, storage pouch
Wireless communication Bluetooth® Smart
Frequency Range: 2.4GHz ISM Band (2400 - 2483.5 MHz)
Modulation: GFSK
Eective radiated power: <20dBm
Service life: 5 years.
Important
Read the operating instructions.
If the device is not used within the specied temperature, humidity and atmospheric
pressure ranges, the technical accuracy of the measurement cannot be guaranteed.
Classication:
Internally powered equipment
Type BF applied parts
Protected against solid foreign objects of 12.5 mm diameter and larger.
Protected against water drops falling vertically when the device is tilted up to 15°.
Not suitable for use in the presence of ammable anaesthetic mixtures with air, oxygen or nitrous oxide.
Continuous operation with short-time loading.
10 °C
40 °C
Operating temperature Storage temperature
Storage humidity/operating humidity
Subject to change without notice.