Specifications
Model: BP 2550
Operation principle: Oscillometric method
Display: Liquid Crystal Display
Measuring range: 0 300 mmHg (cuff pressure)
40 160 / min (pulse rate)
Blood pressure
measurement: 50 mmHg (minimum diastolic value)
250 mmHg (maximum systolic value)
Laboratory accuracy: ± 3 mmHg (cuff pressure)
± 5 % of reading (pulse rate)
Clinical accuracy: According to AAMI-SP 10 with auscultatory reference:
< 5 mmHg systematic offset
< 8 mmHg standard deviation
Inflation: Automatic, adapted to blood pressure
Deflation: Electronic control valve adapted to heart frequency
Power supply: 2 batteries, type LR 03 (AAA)
(e.g. Duracell Ultra)
Operating temperature: + 10 °C to + 40 °C
Storage temperature: 20 °C to + 60 °C
Humidity: Up to 85 % RH
Cuff: Suitable for wrist circumference of 13.5 to 20 cm
Clinical validation: According to AAMI-SP10 with auscultatory reference.
(Please note that a validation against intra-arterial
measurements can lead to different results for devices
which are calibrated against auscultatory reference.)
6063010_BP2550_S4-58 Seite 12 Dienstag, 8. April 2003 2:17 14

Other manuals for Braun SensorControl BP 2550

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Braun SensorControl BP 2550 and is the answer not in the manual?

Braun SensorControl BP 2550 Specifications

General IconGeneral
BrandBraun
ModelSensorControl BP 2550
CategoryBlood Pressure Monitor
LanguageEnglish

Related product manuals