3. Gently pull o the disposable Hygiene cap (2).
Do not use force in removing the
disposable Hygiene cap.
4. Put on a new, clean, disposable Hygiene
cap by snapping into position.
5. Place the protective cap on (1).
The Braun ThermoScan Ear thermometer
turns o automatically after 60 seconds of
inactivity. The thermometer can be turned o
by pressing and holding the start button for
more than 5 seconds.
Disposable Hygiene caps
Use a clean disposable Hygiene cap with each measurement:
1. To ensure hygienic use
2. To protect your child from cross-contamination
3. To ensure earwax residues do not impact accuracy of the
reading.
Earwax and dirt on the tip of the disposable Hygiene cap, even
in microscopic layers, can create an obstacle between the
sensor and the infrared heat emitted by the ear and impact
the accuracy of the measurement.
Temperature-taking hints
Replace the disposable Hygiene cap with every use to to
maintain accuracy, hygiene and to prevent cross-
contamination.
A measurement taken in the right ear may dier from the
measurement taken in the left ear. Therefore, always take
the temperature in the same ear.
The ear must be free from obstructions or excess earwax
build-up to take an accurate reading.
External factors may inuence ear temperatures, including
when an individual has:
– been lying on one ear or the other
had their ears covered
been exposed to very hot or very cold temperatures, or
been recently swimming or bathing.
In these cases, remove the individual from the situation
and wait 20 minutes prior to taking a temperature.
If ear drops or other ear medications have been placed in
the ear canal of one ear, take the temperature in the
untreated ear.
English

Table of Contents

Other manuals for Braun thermoscan IRT 3030

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Braun thermoscan IRT 3030 and is the answer not in the manual?

Braun thermoscan IRT 3030 Specifications

General IconGeneral
Measurement MethodInfrared
Measurement Time1 second
Memory FunctionLast reading
DisplayLCD
Battery Type2 AAA batteries
Automatic Shut-offYes
TypeEar thermometer
Measurement SiteEar
Age RangeSuitable for all ages
HygieneDisposable lens filters
Accuracy±0.2°C (±0.4°F)