Page #10 background image
10
Trouble-shooting
Error message Situation Solution
No lens filter is attached. Attach new, clean lens filter.
The thermometer cannot identify
a secure position of the probe.
An accurate measurement was
not possible.
POS = position error
IRT 4020:
Clear the display by
pushing the «start» button once.
IRT 4520: Clear the display by
pushing the «I/O» button once.
Take care that the positioning of
the probe is correct and remains
stable.
Ambient temperature is not within
the allowed operating range
(10 40 °C or 50 104 °F).
Allow the thermometer to remain
in a room for 30 minutes where
the temperature is between
10 and 40 °C or 50 and 104 °F.
Temperature taken is not within
typical human temperature range
(34 42.2 °C or 93.2108 °F).
HI = too high
LO = too low
Make sure a new, clean lens filter
is attached and thermometer is
properly inserted. Then, take a
new temperature.
System error – self-check display
flashes continuously and will not
be followed by the ready beep
and the ready symbol.
If error persists,
If error still persists,
Wait 1 minute until the thermo-
meter turns off automatically, then
turn on again.
... reset the thermometer by re-
moving the batteries and putting
them back in.
… have the thermometer checked
at a Braun Customer Service
Center.
6022435_IRT_Korea_S4-30 Seite 10 Donnerstag, 24. August 2006 2:30 14

Other manuals for Braun ThermoScan IRT 4020

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Braun ThermoScan IRT 4020 and is the answer not in the manual?

Braun ThermoScan IRT 4020 Specifications

General IconGeneral
BrandBraun
ModelThermoScan IRT 4020
CategoryThermometer
LanguageEnglish

Related product manuals