Model: IRT4520/4020
Displayed temperature range: 34 – 42.2 °C (93.2 – 108 °F)
Operating ambient temperature range: 10 – 40 °C (50 – 104 °F)
Display resolution: 0.1 °C or °F
Accuracy for displayed temperature range: ± 0.2 °C (35.5 – 42 °C) (95.9 – 107.6 °F)
± 0.3 °C (outside this temperature range)
Clinical repeatability: ± 0.14 °C (± 0.26 °F)
Battery life: 2 years / 1000 measurements
Type BF equipment Operating temperature
See instruction of use
Subject to change without notice.
This appliance conforms to the following standards:
Standard Reference Edition Title:
EN 12470-3 2000 Clinical thermometers - Part 3: Performance of compact
electrical thermometers (non-predictive and predictive)
with maximum device
EN 60601-1 2006 Medical electrical equipment – Part 1: General
requirements for basic safety and essential
performance.
EN ISO 14971 2007 Medical devices – Application of risk management to
medical devices.
EN ISO 10993-1 2003 Biological evaluation of medical devices – Part 1:
Evaluation and Testing.
EN 60601-1-2 2007 Medical electrical equipment – part 1-2: General
requirements for basic safety and essential performance
– Collateral standard: electromagnetic compatibility
– Requirements and tests
EN 980 2008 Symbols for use in labeling of medical devices
EN 1041 2008 Information supplied by the manufacturer of medical
This product conforms to the provisions of the EC directive 93/42/EEC
(Medical device directive).
Medical electrical equipment needs special precautions regarding EMC. For detailed description
of EMC requirements please contact your authorized service centre.
Hong Kong service center contact number: +852 2686 3319
Taiwan service center contact number: +886 0800278789
China service center contact number: +86 021-61280577
Singapore service center contact number: +65 65522422
Portable and mobile RF communications equipment can affect medical electrical equipment.
Please do not dispose of the product in the household waste at the end of it useful life.
To protect the environment, dispose of empty batteries at appropriate collection sites
according to national or local regulations.