Page #10 background imageLoading...
Page #10 background image
8
Replacing the battery
The thermometer is supplied with one 3V Lithium cell battery
(type 2032). Insert new battery when the low battery symbol
appears on the display.
The battery door design is
intended to prevent a small
child from easily opening the
battery door to address a
possible choke hazard. Open
the battery compartment by
inserting a narrow pointed object such as a ball point pen or
paperclip into the small hole above the battery door (8). While
pressing into the hole slowly slide back the battery door with
the other hand.
To close battery door align the tabs along the side of the
battery into the battery compartment and slowly slide back
into position until the door clicks into its locked position.
Calibration
This product has been calibrated at the time of manufacture.
If used according to the use instructions, periodic calibration is
not required.
Manufacturing date is given by the LOT number located in the
battery compartment.
The rst three (3) digits after LOT represent the Julian date that
the product was manufactured and the next two (2) digits
represent the last two numbers of the calendar year the
product was manufactured. The last identiers are the letters
that represent the manufacturer.
An example: LOT 11616fam, this product has been
manufactured on the 116th day of the year 2016.