9
English
Product specifications
Model: IRT3030LA
Displayed temperature
range: 34 °C – 42.2 °C
Operating ambient
temperature range: 10 °C – 40 °C
Storage temperature: -25 °C – 55 °C
Display resolution: 0.1 °C
Accuracy for displayed
temperature range: ± 0.2 °C (35.5 – 42 °C)
± 0.3 °C (outside this temperature range)
Clinical bias: -0.11 °C
Limits of agreement:
± 0.25 °C
Clinical repeatability: ± 0.10 °C
Measuring Site: Ear
Reference Site: Oral (estimated oral temperature is
displayed)
Operating and storage
relative humidity: ≤ 95% non-condensing
Battery life: 2 years /1000 measurements
Approximate Service life: 5 years
Type of battery: Lithium 2032
If device is not used or stored within specied temperature and
humidity ranges the technical accuracy cannot be ensured.
Type BF
equipment
See instructions
for use
Operating
temperature
Storage
temperature