17Vessel Distribution
Vessel Distribution
Vessel Distribution
Even vessel distribution is required for optimal performance. Proper spacing will eliminate hot spots, issues
regarding turntable balance, and provide optimal temperature sensing.
Number of Vessels Vessel Distribution
1 - 16 Inner row only - even distribution
17 - 24 Outer row only - even distribution
25 - 40 Inner and outer rows - even distribution
CAUTION
A minimum of 4 vessels per run is required. An error message will appear if <4 vessels are detected.
Vessel distribution is not shown for the 24 place turntable. Ensure vessels are evenly distributed.
The images below are for reference. Not all vessel congurations are displayed.