18 Methods
Methods
One Touch Methods supplied with a MARS 6 are methods created by CEM Applications Specialists. This takes the
guesswork out of method creation and sample preparation. With these methods, one touch (selecting the method)
determines all of the parameters, adjusts power output, and performs the digestion for all major sample types.
Visit our website for Method and Application Notes or contact CEM Corporation before creating a method. Creation
of a method for use in the MARS 6 requires training and knowledge.
To view method parameters, select the method and select the Information icon located at the top of the screen.
Create a One Touch Method
If there is not a current One Touch Method to digest a sample, a One Touch method can be programmed as
outlined below.
1. From the MARS 6 home screen, select “One Touch Methods.”
2. Select “+” at the top of the screen to begin creating a One Touch Method.
3. Select “Method Name” and enter in a name.
4. Select “Temperature” and enter the desired temperature.
5. Select "Sample Type” and choose the appropriate sample type from the list.
• “Organic” is the recommend sample type for most samples. Contact CEM before selecting any other sample
types.
6. Select “Hold Time” and enter in the appropriate hold time.
7. If applicable, select “Sample Prep Notes.”
8. The MARS 6 will complete all other necessary information to perform the method.
9. Select “Next.” View the method by scrolling down. If any parameters require revisions, select “Previous” and
make appropriate revisions.
10. Once the method parameters are entered properly, select the Save icon located in the top right corner.
11. Once the method is saved, the method may be run by selecting the Play icon.
CAUTION
Observe the entire digestion sequence when establishing a procedure for the rst time.
Create a Classic Method
Contact CEM Corporation before creating Classic Methods.