OPERATION LBB Series Oven Owner’s Manual
34 Version 20
Expressed in degrees, from –199 to 999. Use this
parameter to calibrate oven temperature.
Provides levels of access to the controller:
• 0 = No key lock, full access to controller
• 1 = Prevents changing of all parameters except
setpoint
• 2 = Prevents all parameters from being changed
including the setpoint
• -1 = Set to enter the Set-up parameter setting display.
Only when I & d are OFF. Expressed in percent.
Controller outputs this value when process variable equals
setpoint, but only if I = 0.
Only when CtL is OFF, change CtL to PID)
Expressed in degrees. When CtL = OFF, this value
determines the change in temperature needed to turn
controller output from full OFF to full ON.
*If P is not displayed the Control Mode (CtL) must be first set to PID.
5.4.2.2. Control Instrument Operating Mode Notes
• The controller is factory set to LoC=1 (allowing only the setpoint to be changed). This setting
prevents inadvertent changing of control parameters.
• If necessary to alter control parameters, change the LoC=1 to LoC=0. See Section 5.4.2.1 for
more information.
• When LoC = -1, parameters are displayed in the order shown in the Set-Up Parameters table
(Table 5).
• Note: If unable to change setup parameters:
o Change the LoC = 0
o Press the
until LoC appears again
o Change LoC = -1
o Press the
Table 5. Control Instrument Configuration Parameters.
Set for type of input and whether PV is in °C or °F
The maximum setpoint limit for oven. The user cannot
set the setpoint above the maximum setpoint.
The minimum setpoint limit of oven. The user cannot
change the setpoint below this lower setpoint limit.
Despatch recommends leaving the SC feature OFF
Direct/Reverse action. This value is set for 0 (reverse
action).
Priority of PV/SP display. This parameter either
displays the PV (process variable) or SP (setpoint).
• 0 = process variable
• 1 = setpoint