16 Dräger X-cal 2000
Maintenance
5 Maintenance
5.1 Cleaning
1. Disconnect the X-cal 2000 breath alcohol simulator power before cleaning.
2. Do not submerge the top housing in water.
3. Clean the top housing by wiping with a water dampened cloth.
4. Wash the solution container with pla
in water and dry with paper towels.
6 Disposal
7 Technical data
Height: 200 mm
Diameter: 105 mm
Weight: 1.1 kg (with power supply but w/o solution)
Operating temperature: 23 °C ± 5 °C
Storage temperature: 5 to 40 °C
Storage humidity: 10 to 85% RH
Warm-up time from cold start: < 15 minutes
Recovery time from test: < 5 seconds
Power supply rating: 24 VDC, 2.5 A, works with both 110 V and 220 V supplies
8 Order list
CAUTION
Risk of burns!
After disconnecting the X-cal 2000 simulator power, allow 10 to 15 minutes for the
heating element to cool off before detaching the top housing.
This product must not be disposed of as household waste. This is indicated by the ad-
jacent symbol. You can return this product to Dräger free of charge. For information
please contact the national marketing organisations or Dräger.
Name and description Order No.
Dräger X-cal 2000 complete 83 26 000
Ethanol standard solution (0.476 mg/L); 0.5 L 67 28 838
Ethanol solution certified (0.400 mg/L); 5.5 L 36 03 386
Ethanol solution certified (0.250 mg/L); 0.5 L 36 02 949
Calibration adapter (100 pcs) 68 05 473
Calibration sticker (100 pcs) 68 05 739
!