bibag V2.0 Technician’s Manual
P/N 490188 Rev. D
Annual Maintenance
Annual bibag maintenance consists of the following:
Perform bibag Inlet Filter Replacement.
Perform bibag Connector Maintenance.
Perform the bibag Pressure Transducers calibration. (see page 12)
bibag Inlet Filter Replacement
Annually replace the filter (P/N M30225) in tubing assembly labeled 1 connected
between the bibag connector and the bibag air separator.
Figure 5 bibag Inlet Filter (M30225)
bibag Connector Maintenance
Annually replace the two (2) o-rings (P/N 640919) on the bibag connector door.
Figure 6 bibag Connector Door O-Rings (640919)
From bibag
connector
To bibag air
separator
Note the direction of the flow
arrow on the filter housing