4400-0024 enFlow Operator’s Manual EN Rev. T 05/12 Page 19 of 26
Appendix A: Technical Specifications
Size
Warmer: 12.7 cm L x 6.6 cm W x 3.0 cm H, (5.0 in. L x 2.6 in.
W x 1.2 in. H)
Controller: 23.6 cm L x 16.8 cm W x 9.7 cm H, (9.3 in. L x
6.6 in. W x 3.8 in. H)
Disposable Cartridge: 11.4 cm L x 3.8 cm W x 1.0 cm H, (4.5
in. L x 1.5 in. W x 0.4 in. H)
Extension Set: 120 mm L x 10.6 mm W,
(4.7 in. L x 0.4 in. W)
Weight
Warmer: (w/o Disposable): 279 g, (9.8 oz.)
Controller: 1.8 kg, (3.9 lb.)
Disposable Cartridge: 33 g (1.2 oz.)
Extension Set: 2 g (0.07 oz.)
Disposable Cartridge and (optional IV Extension Set):
Priming Volume
Disposable Cartridge: 4 mL
(optional IV Extension Set): 0.5 mL
Sterility
Gamma Sterilized
Biocompatibility
ISO 10993
Infusion Set Compatible
ISO 8536-4
Performance
Fluid Temperature Output
40 °C ± 2 °C
Flow Rate Range
KVO to 200 mL/min
Input Voltage
Warmer: 28 VDC at a maximum of 300 Watts
Controller: 110-120 or 220-240 VAC
Temperature Set Point
40 ºC
Over Temperature Set Point
ASTM F-2172-02
Alarms
IEC60601-1-8:2006
Input Current
5 A
Input Frequency Range
Warmer: DC
Controller: 47-63 Hz
Environmental/ Physical Requirements
Temperature, Operating
-5 °C to 50 °C
Temperature, Storage
-30 °C to 70 °C
Water Resistance
Warmer: IEC 529 IP67 30 minutes immersion at a depth of
91.4 cm (36 in.); Controller: IEC 529 IP21 dripping water;
Disposable Cartridge and (optional IV Extension Set): IEC
529 IP68 IV Extension Set): IEC 529 IP68 continuous
immersion
Penetration
Warmer: IEC 529 IP67 dust tight
Controller: IEC 529 IP21 ≥ 12.5 diameter (See Page 23)
Disposable Cartridge and (optional IV Extension Set): IEC
529 IP68 dust IV Extension Set): IEC 529 IP68 dust tight
Electrical Safety
UL 60601-1:2005 R6.03, CAN/CSA-C22.2 No. 60601.1:2008,
IEC 60601-1-6:2010, AAMI ES60601-1:2005, IEC 60601-1-
4:2000 (Canada)
Relative Humidity, Operating and
Storage
Warmer: 10 % to 90 %
Controller: 10 % to 90 %
Disposable Cartridge and (optional
IV Extension Set): 10 % to 90 %