EasyManuals Logo
Home>GE>Medical Equipment>LOGIQ P6

GE LOGIQ P6 Technical Publication

GE LOGIQ P6
127 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #64 background imageLoading...
Page #64 background image
Site Preparations
3-12 Ultrasound SystemCommon Service Information
Direction 5444964-100 English
Rev. 5
Electrical requirements for Laptop Ultrasound Systems (continued)
Table 3-8: Electrical Requirements - Vivid Laptops
Product Voltage Tolerance
Current or
Power
Consumption
Frequency
AC/DC Converter for
Vivid i/Vivid iN and
Vivid q/ Vivid qN
100 - 240 VAC ±10% 0.5 to 1.0 A 50-60 Hz
Adapter for Vivid e 100 - 240 VAC ±10% 130 VA max. 1.3 A
(max.)
50/60 Hz
Table 3-9: Electrical Requirements - Voluson Laptops
Product Voltage Tolerance
Current or
Power
Consumption Frequency
Voluson i/e 100 - 240 VAC
±10%
120 VA
50, 60 Hz
(±2%)
Voluson Station /
Voluson Dock Cart
100-130 VAC
220-240 VAC 320 VA