General Ultrasound system requirements
Ultrasound SystemCommon Service Information 3-11
Direction 5444964-100 English Rev. 5
Electrical requirements for Laptop Ultrasound Systems
Contents:
• ‘Electrical Requirements - LOGIQ Laptops’ on page 3-11
‘Electrical Requirements - Vivid Laptops’ on page 3-12
‘Electrical Requirements - Voluson Laptops’ on page 3-12
‘Docking Cart for LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid| e/
Vivid iq Power Requirements’ on page 3-13
• ‘Isolation Cart for LOGIQBook XP Series/LOGIQ e/
LOGIQ e Vet/LOGIQ i/Vivid e’ on page 3-14
Table 3-7: Electrical Requirements - LOGIQ Laptops
Product Voltage Tolerance
Current or
Power
Consumption
Frequency
LBAC-66 adapter for
LOGIQ Book
100-240 VAC ±10%
120 VA 1.2-0.5 A
50-60 Hz
GE-90W adapter for
LOGIQ Book
108 VA 1.08-0.45
A
ADM-9020M-GE for
LOGIQ Book XP Series
Adapter for LOGIQ e and
LOGIQ e Vet R7.x.x and
below.
130 VA max. 1.3 A
(max.) 50/60 Hz
Adapter for LOGIQ i
Adapter for LOGIQ e
R8.x.x, R9.x.x
160 VA max. 1.6 A
(max.)
AC Adapter for LOGIQ
V2/V1
2 A 50-60 Hz

Table of Contents

Other manuals for GE LOGIQ P6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P6 and is the answer not in the manual?

GE LOGIQ P6 Specifications

General IconGeneral
BrandGE
ModelLOGIQ P6
CategoryMedical Equipment
LanguageEnglish

Related product manuals