Loading the software
LOGIQ V3/V5/V5 Expert – Basic Service Manual 8-19
5726264-100 English Rev.8
8-3-7-3 Loading the system with USB memory stick(continued)
9. After finish updating system, remove the USB memory stick
and press any key to shut down the system.
NOTE: If you do not remove the USB memory stick, the software
system loading process repeats when the system boots up.
Do not remove the USB stick after the software reload if you
want to adjust the screen size.
Figure 8-13. System upgrade complete
10. Password, serial number and option key are needed to be
input after loading the complete disk.
NOTE: The serial number and option key MUST be the same as the
ones before installation. Please record the serial number
and option key before installation.