Facility needs
Ultrasound SystemCommon Service Information 3-25
Direction 5444964-100 English Rev. 5
Required facility needs (continued)
Dedicated single branch power outlet of adequate
amperage, meeting all local and national codes, which is
located less than 2.5 m (8 ft.) from the Ultrasound system’s
proposed location
Door opening is at least 76 cm (30 in) wide
Proposed location for Ultrasound system is at least 0.5 m
(1.5 ft.) from the wall for cooling
Power outlet and place for any external peripheral are within
2 m (6.5 ft.) of each other with peripheral within 1 m of the
Ultrasound system to connect cables.
Power outlets for other medical equipment
Power outlets for test equipment within 1 m (3.2 ft.) of
Ultrasound system
Clean and protected space to store probes (in their cases or
on a rack)
Material to safely clean probes (done with a plastic
container, never metal)
For the amperage requirements, see: ‘Electrical requirements’
on page 3-5.
Desirable features
Door is at least 92 cm (3 ft.) wide
Circuit breaker for dedicated power outlet is easily
accessible
Sink with hot and cold water
Receptacle for bio–hazardous waste, like used probe
sheaths
• Emergency oxygen supply
Storage for linens and equipment
Nearby waiting room, lavatory, and dressing room
Dual level lighting (bright and dim)
Lockable cabinet ordered by GE for its software and
proprietary manuals