MX-7711-02
2. Preparations for use
For more detailed information see section 3 in this letter.
Check that the manual resuscitator is readily available and is working correctly.
Remove the vaporizers by lifting up the front and gently pulling outwards.
If the system is equipped with N
2
O gas module and/or with the alarm options Mute Apnea Alarms
and/or HLM/CPB (heart lung machine mode/cardio
pulmonary bypass) these should be turned off in the
Service & Settings menu. The alarm modes above will
silence and, for HLM/CPB, disable crucial alarms and
are not to be used in the ICU. Please contact a certified
Getinge trained service engineer to assist.
Connect breathing circuit according to user manual.
Use a high quality antiviral HMEF (heat and moisture
exchange filter) placed between the breathing circuit and the patient’s airway, this will protect the
machine from contamination.
The gas sampling line should be connected to the HMEF on the machine side of the filter to ensure
sampled gas is not contaminated as it enters the water trap.
Make sure to use an expiratory bacterial/viral filter, to minimize cross-contamination. Make sure to
follow the filter manufacturer’s recommendations for exchange interval of the filters.
Perform a system checkout (SCO) according to instructions on screen.The test will take 6-8 minutes.
Back-up gas cylinders are included in the check out procedure should they be connected.
Set alarm sound level to 100% and modify alarm limits accordingly