8 TROUBLESHOOTING
WARNING: If you experience a problem with your lift and are unable to service it
yourself, contact your Lumex dealer.
LF1031 HYDRAULIC PATIENT LIFT
SYMPTOM POSSIBLECAUSE ACTION
Patientliftdoesnotgoup
ordown
Airinthecylinder Bleedthecylinderbyturningtherelease
knobandpushdownonthemast.Pump
uptheliftagainandturnthereleaseknob
again.
Lossofhydraulicuid(leaking
cylinder)
Replaceanyleakingcylinder
Patientliftdoesnotgo
down
Airinthecylinder SeeaboveDirections
Info: If the hydraulic cylinder begins leaking fluid, do not operate lift. Contact your Lumex
dealer for a new cylinder.
LF1041 BATTERY POWERED PATIENT LIFT
SYMPTOM POSSIBLECAUSE ACTION
Controlboxisnot
respondingandrelayisnot
clicking
Batterydischarged Chargebattery
Batterychargerdefective Replacebatterycharger
Batterydefective Replacebattery
Handcontrolpendantdefective Replacehandcontrolpendant
Actuator/motordoesnot
run,butrelayisclicking
Actuatorplugnotproperlyseated Pushactuatorplugrmlyintocontrolbox
Actuatordefective Replaceactuator
Controlboxdefective Replacecontrolbox
Actuator/motorgoesup
ordown,butnotboth
Handcontrolpendantdefective Replacehandcontrolpendant
Controlboxdefective Replacecontrolbox