6.3.6 Interval Setup
A setup created from the control panel will
assign all channels to the same logging
interval. To change or view this interval choose
‘Intervals’ from the main ‘Setup’ menu.
6.3.6 Sensor Power Setup
The sensor power setup can be changed by
going to ‘Sensor Power’ from within the main
‘Setup’ menu.
Important Notes on Setup using the Control Panel
Whilst all channels are assigned to the same interval, Interval A, all intervals can be
viewed in the case of a more complex setup from SquirrelView
To avoid problems with wiring configuration, sensor types are not editable once set but
their range and sensor power requirements can be. Delete a channel if you require to
change its Sensor Type
If the number of sensors is exceeded when adding new channels, the message "Invalid
Setup" will be displayed
Calculated channels setup from SquirrelView will not be shown on the view channels
option
The reference junction can not be deleted while a sensor type that requires it for
correct operation is set up, e.g. a thermocouple
You cannot change a setup whilst the logger is armed
The message "Sub-Second rate" will be displayed when attempting to view or edit a
Sample Intervals that have been set to less than 1 second. Use Squirrelview Logger
Setup to view or edit the sub-second Intervals
Full setup capabilities of the 2020/2040 are provided using the SquirrelView software
provide. When using the control panel a few important points should noted:-