JB Academy, JB Nova and SUB Aqua Pro 31394 V7
Detailed operating manual Page 21
www.grantinstruments.com
Firstly set the bath to the lower working temperature and allow to stabilise for at least an
hour.
Place the reference thermometer either in the centre of the bath, or if using a lid, through
the thermometer hole.
Note the value of the reference thermometer and enter it into the calibration menu by
following the steps below.
Figure 8 - Setting the low point calibration
Select the menu
option
Select the low
point calibration
Press set to enter
the calibration value
Enter the reference
thermometer value
Press set to store
the calibration value
Leave the bath to
stabilise and check
the calibration



.
.