EasyManuals Logo
Home>Grant>Laboratory Equipment>SAP2

Grant SAP2 User Manual

Default Icon
28 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #22 background imageLoading...
Page #22 background image
JB Academy, JB Nova and SUB Aqua Pro 31394 V7
Detailed operating manual Page 22
www.grantinstruments.com
Secondly set the bath to the upper working temperature and allow to stabilise for at least
an hour.
Place the reference thermometer either in the centre of the bath, or if using a lid, through
the thermometer hole.
Note the value of the reference thermometer and enter it into the calibration menu by
following the steps below.
.
Figure 9 - Setting the high point calibration
Select the menu
option
Select the high
point calibration
Press set to enter
the calibration value
Enter the reference
thermometer value
Press set to store the
calibration value
Leave the bath to
stabilise and check
the calibration


.
.


Table of Contents

Other manuals for Grant SAP2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Grant SAP2 and is the answer not in the manual?

Grant SAP2 Specifications

General IconGeneral
BrandGrant
ModelSAP2
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals