MPD Operating Manual
27
Sensor
Type
Drift Over
Time
Operating Temperature
Operating
Pressure
Operating
Air Speed
Min Max
MPD-IC1 < 3%/yr -20°C / -4°F +50°C / 122°F 80kPa ~ 110kPa 0 ~ 6m/sec
MPD-IV1 < 3%/yr -20°C / -4°F +50°C / 122°F 80kPa ~ 110kPa 0 ~ 6m/sec
MPD-IF1 < 3%/yr -20°C / -4°F +50°C / 122°F 80kPa ~ 110kPa 0 ~ 6m/sec
MPD-CB1 < 3%/yr -40°C / -40°F +65°C / 149°F 80kPa ~ 120kPa 0 ~ 6m/sec
NOTeS
• Data taken at 20-25°C. Contact HA for any additional data or details.
• Response times may increase at lower temperatures.
• Data represents typical values without optional accessories attached
(MPD**IC1 with SPXCDWP. T5<60, T90<120).
• Note: Environmental conditions that compromise the IP66 protection provided
by the weather proof cover will extend published response times. Safety
protocols or maintenance procedures that consider these environmental
conditions are recommended specific to the installation.
• System conditioning may be required to achieve stated results. Contact HA
for details.
• Performance figures are measured using a sample humidity of 50% RH.
• Performance figures are measured between 40 and 60% of full scale.
• Performance figures are measured by test units calibrated at 50% of full scale.